Experimental validation of motion sensor Physilog®5 applied to shoulder joint

Authors

  • Jan Krivošej Czech Technical University in Prague, Faculty of Mechanical Engineering, The Department of Mechanics, Biomechanics and Mechatronics, Technická 4, 160 00 Prague 6, Czech Republic
  • Jana Garanová Krišťáková Czech Technical University in Prague, Faculty of Mechanical Engineering, The Department of Mechanics, Biomechanics and Mechatronics, Technická 4, 160 00 Prague 6, Czech Republic
  • Matej Daniel Czech Technical University in Prague, Faculty of Mechanical Engineering, The Department of Mechanics, Biomechanics and Mechatronics, Technická 4, 160 00 Prague 6, Czech Republic
  • Zbyněk Šika Czech Technical University in Prague, Faculty of Mechanical Engineering, The Department of Mechanics, Biomechanics and Mechatronics, Technická 4, 160 00 Prague 6, Czech Republic

DOI:

https://doi.org/10.14311/APP.2023.42.0055

Keywords:

shoulder joint, motion sensor, wearable, experimental validation

Abstract

The main motivation of this paper is to verify the idea of using the Physilog®5 unit for the patients with shoulder movement difficulties. The attached sensor to the patient’s arm then measures motion during which the patient should follow certain paths. Finally, if a patient has difficulty with motion requirements, some typical pattern for their problem should emerge. By analysing these patterns, a database of typical problems could be created, which could assist doctors in determining a patient’s diagnosis.The experiment is focused on Physilog® concerning the 5th generation. The goal is to experimentally identify and verify the performance of this generation during relatively large motions of the upper limb. For this purpose, an experimental stand representing spherical joint with an accurate absolute position sensing is assembled and calibrated. Subsequently, the three Physilog®5 sensors are mounted on this stand at different positions.

Downloads

Download data is not yet available.

Downloads

Published

2023-10-12

How to Cite

Krivošej, J., Garanová Krišťáková, J., Daniel, M., & Šika, Z. (2023). Experimental validation of motion sensor Physilog®5 applied to shoulder joint. Acta Polytechnica CTU Proceedings, 42, 55–60. https://doi.org/10.14311/APP.2023.42.0055